Chlorine Reagent Tablets CL203
Chlorine Reagent Tablets CL203
เม็ดน้ำยาสำหรับวัดค่า คลอรีน
Features:
- Includes 10 packs (10 tablets ea.)
- Requires 1 tablet per test - enough for 100 tests
- Foil packaging provides very stable environment with long shelf life
- No messy powder reagents
- No dosing equipment required
Instructions for use
- Fill a sample cup with exactly 20ml of the test solution (up to the mark on the cup).
- Drop one (1) ExTab® reagent tablet into sample cup. Do not handle the reagent tablet with fingers.
- Let tablet soften for 10 seconds then shake for 20 seconds (or until tablet is dissolved).
- Firmly wipe the electrode tip with a dry paper towel to remove any compounds that may have formed.
- Turn the CL200 on and immediately immerse in the solution.
- Swirl in solution for 15 seconds.
- The display will flash as long as the readings are changing rapidly. Read the total chlorine content when the display “holds” the reading.
- Do not turn the ExStik® off while in the solution
- Remove the CL200 from the sample solution and rinse the electrode with de-ionized water.
- Wipe off the electrode with a paper towel and turn it off.
- Replace protective cap.